Summary
A new custom imaging system was developed for biological assay development, enabling fluorescent macroimaging of large agar plates up to 245 x 245 mm with selectable wavelengths.
Area of Science:
- Biological assay development
- Imaging technology
- Biotechnology
Background:
- Advancements in biological assay development necessitate improved imaging technologies.
- Existing commercial systems lack the capability for fluorescent macroimaging of large agar plates with discrete wavelength selection.
Purpose of the Study:
- To address the limitations of current commercial imaging systems.
- To design and construct a custom imaging system for macroimaging biological samples.
Main Methods:
- A custom imaging system was designed and constructed.
- The system was engineered to capture both white light and fluorescent images.
- It accommodates agar plates from NUNC Omni tray (96-well footprint, 128 x 85 mm) to NUNC Bio Assay Dish (245 x 245 mm).
Main Results:
- The custom imaging system successfully provides fluorescent macroimaging capabilities.
- It supports imaging areas up to 245 x 245 mm.
- The system allows for discrete wavelength selection within the visible light spectrum.
Conclusions:
- The developed custom imaging system meets the technological requirements for advanced biological assay development.
- This system offers a viable solution for researchers needing to fluorescently image large agar plates with wavelength specificity.
Related Concept Videos
Confocal Fluorescence Microscopy
Confocal microscopy is an advanced microscopic technique. The prime advantage of the confocal microscope over other microscopy techniques is its ability to block the out-of-focus light from the illuminated samples using pinholes. It is widely used with fluorescence optics to obtain high-resolution, sharp contrast images. Unlike optical microscopes, confocal microscopes use a focused beam of light laser to scan the entire sample surface at different z-planes. These microscopes are, therefore,...
Super-resolution Fluorescence Microscopy
Super-resolution fluorescence microscopy (SRFM) provides a better resolution than conventional fluorescence microscopy by reducing the point spread function (PSF). PSF is the light intensity distribution from a point that causes it to appear blurred. Due to PSF, each fluorescing point appears bigger than its actual size, and it is the PSF interference of nearby fluorophores that causes the blurred image. Various approaches to achieving higher resolution through SRFM have recently been developed.