Importance of dosimeter calibration method on nebulizer output
David C Todd1, Beth E Davis, Audrey J Smycniuk
1Division of Respiratory Medicine, Department of Medicine, University of Saskatchewan, Royal University Hospital, Saskatoon, Canada.
Background:
We have observed that dosimeter-run nebulizers have a much smaller output when manually activated than when breath activated; however, this has not been adequately investigated.
Objective:
To evaluate the effect of different calibration methods on nebulizer output.
Methods:
Six healthy subjects performed all calibrations. The nebulizers were operated by 2 different dosimeters and were calibrated to produce 9 microL per actuation by breath activation followed by exhalation to the room. The nebulizers were then operated at these identical settings, and the output determined in 3 ways: (1) breath activation followed by exhalation to the room, (2) breath activation with exhalation into the nebulizer, and (3) manual activation (with no subject using the nebulizer). These 3 methods were termed regular, rebreathe, and manual, respectively.
Results:
There was a large and statistically significant difference in nebulizer output among the 3 methods. The measured rebreathe outputs (5.6 and 5.7 microL per actuation) were approximately two thirds and the manual outputs (3.2 and 3.9 microL per actuation) were approximately one third of the regular calibration outputs (8.6 and 8.9 microL per actuation); the 2 values are for the 2 dosimeters. The results were highly statistically significant (P < .001).
Conclusions:
The method by which a nebulizer-dosimeter system is calibrated results in different nebulizer outputs. This has a high likelihood of influencing the concentration of methacholine causing a 20% decrease in volume in the first second of forced expiration.
More Related Videos
08:59Measuring Sub-23 Nanometer Real Driving Particle Number Emissions Using the Portable DownToTen Sampling System
Published on: May 22, 2020
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Flame Photometry: Lab
Inhaled Medications
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...