Related Experiment Video
Updated: Aug 11, 2026

Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
[Gas chromatographic determination of residual toluene in plastic packing bags]
1National Chromatographic R. & A. Center, Dalian Institute of Chemical Physics, the Chinese Academy of Sciences, Dalian, 116011.
Abstract:
A method for analysis of residual toluene in plastic packing bags by gas chromatographic headspace analysis is presented. This method is very simple and accurate. It only requires one column: 0.5 m x 2 mm i.d. packed with 102 white support DMCS coated with 20% PEG-20M. The relation between the preheating temperature of sample and the peak height of toluene can be approximately expressed as H=4.156e(0.126T). The precision of this method is 4% and its detectable limit is 0.031 mg (toluene)/m2.
Related Concept Videos
Volatilization
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Microbial Bioremediation of Plastics