Related Experiment Video
Updated: Aug 18, 2026

Sample Preparation in Quartz Crystal Microbalance Measurements of Protein Adsorption and Polymer Mechanics
Published on: January 22, 2020
Molecularly imprinted ligand-exchange recognition assay of glucose by quartz crystal microbalance
Arzu Ersöz1, Adil Denizli, Ali Ozcan
1Department of Chemistry, Anadolu University, Eskişehir, Turkey.
Abstract:
Molecular imprinted polymers (MIP) as a recognition element for sensors are increasingly of interest and MIP-quartz crystal microbalance (QCM) have started to appear in the literature. In this study, we have combined quartz crystal microbalance with MIP to prepare a sensor using the ability of glucose to chelate of copper (II) ion of methacrylamidohistidine (MAH) monomer to create ligand exchange (LE) assembled monolayer which is suitable for glucose determination. The study includes the measurement of binding interaction of molecularly imprinted QCM sensor via ligand interaction, investigation of the pH effect on frequency shift and recognition selectivity studies of glucose-imprinted polymer with respect to methyl-alpha-d-glucopyranoside and sucrose. Bmax (number of binding sites) and K(D) (dissociation constant of the metal-chelate copolymer) were also calculated using Scathard plot and the detection limit was found as 0.07 mM. MIP showed higher glucose-binding affinity than a well-known glucose binding protein, conconavalin A.