Effects of high-frequency electromagnetic fields emitted from card readers of access control systems on electronic
Shizuhiko Deji1, Kunihide Nishizawa
1Graduate School of Environmental Studies, Nagoya University, Furo-cho, Chikusa-ku, Nagoya 464-8602, Japan.
Abstract:
High-frequency electromagnetic fields in the 120 kHz band emitted from card readers for access control systems caused abnormally high doses on electronic pocket dosimeters (EPDs). All EPDs recovered their normal performance by resetting after the exposure ceased. The electric and magnetic immunity levels of the EPDs were estimated by using the distances needed to prevent electromagnetic interference.
More Related Videos
06:21Diffuse Optical Spectroscopy for the Quantitative Assessment of Acute Ionizing Radiation Induced Skin Toxicity Using a Mouse Model
Published on: May 27, 2016
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Related Concept Videos
Biological Effects of Radiation
Electronic Distance Measuring Instruments
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...