Corneal oxygenation during contact lens wear: comparison of diffusion and EOP-based flux models
1Brennan Consultants, 110 Auburn Road, Auburn Village, VIC, 3122, Australia.
Purpose:
The aim of this study was to compare corneal oxygen flux values derived from an oxygen diffusion model, with estimates from a model in which equivalent oxygen percentage (EOP) values were substituted for the post-lens tear film oxygen tension in Fick's law.
Methods:
A previously described five-layer corneal oxygen diffusion model was found to artefactually allow theoretical oxygen consumption, when the predicted oxygen tension fell to zero. Consequently, an eight-layer diffusion model was constructed, with consumption set to zero at points within the cornea, where predicted oxygen tension falls to zero. Post-lens tear layer thickness was corrected to more contemporary estimates. The eight-layer and EOP-based anterior corneal oxygen flux estimates were compared across the range of commonly encountered contact lens Dk/t values.
Results:
The eight-layer model overcomes deficiencies in the five-layer model and provides predicted values that are remarkably similar to the EOP-based model. Open and closed eye anterior corneal oxygen flux in the absence of contact lens wear was estimated at 7.8 and 7.6 microL/cm2/hr for the open eye and 6.0 and 6.1 microL/cm2/hr for closed eye for the diffusion and EOP-based models, respectively.
Conclusions:
The diffusion model supports the EOP model in that there is minimal oxygen benefit to be gained by increasing Dk/t above the Holden-Mertz criteria of 24 and 87 x 10(-9) (cm/sec)(ml0(2)/ml.mmHg) during open and closed eye wear, respectively. The eight-layer model is suitable for further definition of corneal oxygenation during contact lens wear.
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Physiological Pharmacokinetic Models: Blood Flow-Limited Versus Diffusion-Limited Models
Ophthalmic Drug Delivery Systems