Breast dosimetry using high-resolution voxel phantoms.

D R Dance1, R A Hunt, P R Bakic

  • 1Department of Physics, The Royal Marsden NHS Foundation Trust, Fulham Road, London SW3 6JJ, UK. david.dance@rmh.nthames.nhs.uk

Radiation Protection Dosimetry
|June 4, 2005
PubMed
Summary

Accurate breast cancer screening requires precise radiation dose calculations. This study highlights how the distribution of glandular tissue in mammography significantly impacts dose estimation, necessitating improved models for patient safety.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Image quality assessments in abdominal CT: Relative importance of dose, iterative reconstruction strength and slice thickness.

Radiography (London, England : 1995)·2024
Same author

IMAGE QUALITY AND POTENTIAL DOSE REDUCTION USING ADVANCED MODELED ITERATIVE RECONSTRUCTION (ADMIRE) IN ABDOMINAL CT - A REVIEW.

Radiation protection dosimetry·2021
Same author

Establishing the European diagnostic reference levels for interventional cardiology.

Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB)·2018
Same author

Image processing can cause some malignant soft-tissue lesions to be missed in digital mammography images.

Clinical radiology·2017
Same author

Design and application of a structured phantom for detection performance comparison between breast tomosynthesis and digital mammography.

Physics in medicine and biology·2017
Same author

Can the channelized Hotelling observer including aspects of the human visual system predict human observer performance in mammography?

Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB)·2017

Area of Science:

  • Medical Physics
  • Radiological Imaging
  • Biomedical Engineering

Background:

  • X-ray mammography is a key tool for breast cancer detection.
  • Accurate dosimetry is crucial for patient safety and effective screening.
  • Current dosimetry methods may not fully account for individual breast tissue variations.

Purpose of the Study:

  • To develop and utilize a high-resolution computer model for simulating breast tissue composition.
  • To investigate the impact of glandular tissue distribution on radiation dose conversion factors in mammography.
  • To compare model-derived dose conversion factors with standard tabulations.

Main Methods:

  • Development of quasi-realistic, high-resolution (400 microm voxels) 3D breast phantoms simulating adipose, fibro-glandular tissues, ducts, and skin.

Related Experiment Videos

  • Simulation of X-ray mammography using a Mo/Mo spectrum at 28 kV.
  • Calculation of incident air kerma to mean glandular dose (g) conversion factors for phantoms of varying thickness (40-80 mm) and glandularity.
  • Main Results:

    • The computer model accurately simulates complex 3D breast tissue distributions.
    • Calculated 'g' factors showed significant deviations (up to 43% and 9-59% in further studies) from standard tabulations.
    • These discrepancies are attributed to the spatial distribution of glandular tissue within the breast.

    Conclusions:

    • The spatial distribution of glandular tissue is a critical factor in accurate breast dosimetry.
    • Standard dosimetry tabulations may overestimate or underestimate radiation dose due to neglecting tissue heterogeneity.
    • Advanced computational models incorporating detailed tissue distribution are essential for precise mammography dose estimation and patient safety.