Related Experiment Video
Updated: Aug 10, 2026

Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
Cardiac index measurements during rapid preload changes: a comparison of pulmonary artery thermodilution with
Thomas W Felbinger1, Daniel A Reuter, Holger K Eltzschig
1Department of Anesthesiology, Grosshadern Medical Center, University of Munich, 81377 Munich, Germany. tfelbinger@partners.org
Study Objective:
To compare cardiac index (CI) values obtained by pulmonary artery thermodilution (CIPA), arterial thermodilution (CITD), and arterial pulse contour analysis (CIPC) during rapid fluid administration, as accurate and rapid detection of CI changes is critical during acute preload changes for guiding volume and vasopressor therapy in critically ill patients, and the accuracy of CIPC during acute changes in loading condition is currently unknown.
Design:
Prospective clinical study.
Setting:
Cardiac surgical intensive care unit of a university hospital.
Patients:
Seventeen American Society of Anesthesiologists (ASA) physical status II and III patients, aged 32 to 76 years, with normal left ventricular function during the early postoperative period after elective coronary artery bypass graft surgery.
Measurements:
After baseline determinations of CIPA, CIPC, and CITD were made, fluid loading was performed using 10 mL times body mass index of hydroxyethyl starch 6%. CIPA, CIPC, and CITD were determined, and changes in CI (DeltaCI) were calculated. Fluid load was repeated until no increase in stroke volume index (DeltaSVI <10%) was achieved.
Main Results:
Regression analysis between CIPA/CIPC, CIPA/CITD, and CIPC/CITD revealed r2 = 0.92, r2 = 0.92, and r2 = 0.98. Regression analysis between DeltaCIPA/DeltaCIPC, DeltaCIPA/DeltaCITD, and DeltaCIPC/DeltaCITD revealed r2 = 0.57, r2 = 0.67, and r2 = 0.74, respectively. Bland-Altman analysis was used to determine accuracy and precision of the 3 methods compared. The mean differences (m) and SD between DeltaCIPA/DeltaCIPC, DeltaCIPA/DeltaCITD, and DeltaCIPC/DeltaCITD resulted in m = -1.01%, SD = 6.51%; m = -0.83%, SD = 5.80%; and m = -0.33%, SD = 4.65%, respectively.
Conclusion:
Compared with pulmonary artery thermodilution, arterial pulse contour analysis reflects relative changes in CI during rapid changes of preload with clinically acceptable accuracy.
Related Concept Videos
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls between...
Pulse
The pulse serves as a clinical indicator...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...