Multimodal Cross-Device and Marker-Free Co-Registration of Preclinical Imaging Modalities
Published on: October 27, 2023
Automatic detection of fiducial markers in fluoroscopy images for on-line calibration
Laurence Smith1, Mike Pleasance, Rosalyn Seeton
1The University of British Columbia, Vancouver Canada.
Abstract:
Calibrated C-arm fluoroscopy is used for a variety of medical procedures where objects and anatomical structures need to be located in space. Calibration is often based on imaging a grid of fiducial markers and using the C-arm image's geometrical measurements (radius and center) together with the positions of the markers. An on-line technique is developed to automatically locate the fiducial markers and validated on 97 images. The success rate of the detection algorithm is 96.28% with an average error of 0.46 mm and a standard deviation of 0.32 mm.