Influence of elastic and non-elastic external Dacron mesh support on para-anastomotic hypercompliance in end-to-end
1Division of Vascular and Endovascular Surgery, Johann Wolfgang Goethe University, Frankfurt am Main, Germany.
Objective:
To quantify the influence of elastic and non-elastic external mesh support on para-anastomotic hypercompliance in end-to-end anastomoses (ex vivo).
Materials:
Six end-to-end anastomoses prepared from ovine carotid arteries without mesh support and with external elastic and non-elastic dacron mesh support.
Methods:
Compliance profiles of the anastomised arterial segments were measured successively, in the same anastomotic configuration without mesh support, with external elastic dacron mesh support and with external non-elastic dacron mesh support (randomized order). A pulsatile ex vivo perfusion system using a laser scan micrometer to monitor outer systolic and diastolic diameter was employed.
Results:
Median pre-anastomotic and post-anastomotic hypercompliance without external mesh support were 1.45 and 1.19%/100 mmHg, respectively, above reference compliance. Use of the elastic mesh support significantly reduced the median hypercompliance to 0.68%/100 mmHg (pre-anastomotic) and to 0.34%/100 mmHg (post-anastomotic) above reference compliance. The non-elastic mesh support caused approximately the same significantly reduced median hypercompliance to 0.53%/100 mmHg (pre-anastomotic) and 0.43%/100 mmHg (post-anastomotic).
Conclusions:
Both elastic and non-elastic external mesh support significantly reduced pre- and post-anastomotic hypercompliance.