Related Experiment Video
Updated: Aug 16, 2026

Semi-Targeted Ultra-High-Performance Chromatography Coupled to Mass Spectrometry Analysis of Phenolic Metabolites in Plasma of Elderly Adults
Published on: April 22, 2022
Determination of phenols in waters using micro-pumped multicommutation and spectrophotometric detection: an automated
Eva Ródenas-Torralba1, Angel Morales-Rubio, Miguel de la Guardia
1Department of Analytical Chemistry, University of Valencia, 50th Dr. Moliner St., 46100 Burjassot, Valencia, Spain.
Abstract:
An automated and greener spectrophotometric procedure has been developed for the determination of phenol in water at 700 nm. The method uses the reaction between phenol, sodium nitroprusside, and hydroxylamine hydrochloride in a buffered medium at pH 12.3. The flow manifold comprises four solenoid micro-pumps employed for sample and reagent introduction into the reaction coil and to transport the colored product formed to the detector. The linear dynamic range was 50-3,500 ng mL(-1) (R = 0.99997; n = 6) and the method provided a limit of detection (3sigma) of 13 ng mL(-1). The sampling throughput was estimated to be 65 measurements per hour and the coefficient of variation was 0.5% (n = 10) for a 1.0 microg mL(-1) phenol concentration. Recoveries of 92-105% were obtained for phenol determination in spiked water samples at concentration levels from 50 to 5,000 ng mL(-1). The use of multicommutation reduced the reagent consumption 25-fold, the sample consumption 225-fold, and the waste generation 30-fold compared with the batch procedure. The proposed method is an environmentally friendly alternative to the official 4-aminoantipyrine method since it avoids the use of chloroform.

