Related Experiment Videos
Contamination in HPLC quantified benzaldehyde is from polypropylene microtubes.
Tomer Shemesh1, Connie Karschimkus, Kevin G Rowley
1Menzies School of Health Research, Institute of Advanced Studies, Charles Darwin University, John Mathews Building, Royal Darwin Hospital, Rocklands Drive Tiwi NT 0810, Australia. tomer.shemesh@menzies.edu.au
Summary
Contaminated polypropylene vials skewed semicarbazide-sensitive amine oxidase (SSAO) assay results. Borosilicate glass vials are recommended for accurate aldehyde production measurements in SSAO activity assays.
Area of Science:
- Biochemistry
- Enzymology
- Vascular Biology
Background:
- Semicarbazide-sensitive amine oxidase (SSAO) is a copper-containing enzyme found in vascular smooth muscle cells.
- SSAO deaminates primary amines, producing aldehydes and peroxides, potentially contributing to vascular damage.
- Accurate quantification of SSAO activity is crucial for understanding its role in vascular health.
Purpose of the Study:
- To investigate a source of high background benzaldehyde levels observed during SSAO activity assays.
- To evaluate the impact of different vial materials on assay accuracy.
- To recommend optimal conditions for reliable SSAO activity measurement.
Main Methods:
- SSAO activity was quantified by measuring benzaldehyde production.
- Fluorescent derivatization and High-Performance Liquid Chromatography (HPLC) were employed for analysis.
- Derivatization was performed in both polypropylene and borosilicate glass tubes at 95°C for 45 minutes.
Main Results:
- One batch of polypropylene vials exhibited significant background benzaldehyde, interfering with SSAO assay results.
- Borosilicate glass vials did not show this contamination issue.
- Contamination was not linked to shipping, handling, or assay reagents, suggesting a manufacturing-related issue.
Conclusions:
- Variations in the commercial production of polypropylene vials can introduce contaminants affecting biochemical assays.
- Borosilicate glass vials are a reliable alternative for aldehyde production assays.
- Using borosilicate vials is recommended to ensure accurate SSAO activity measurements.