Abstract:
Ultrasound Doppler is widely used for low measurement in both medicine and industry, having the advantages of being non-invasive and comparatively simple and therefore inexpensive. The technique has not however been used for capillary blood flow measurement, because of the relatively low velocities encountered and because of the presence of strong interfering signals from the encompassing tissue. An ultrasound Doppler system capable of measuring flow velocities of one millimetre per second in the presence of one thousand times stronger interfering signals is described, as well as test results using both thread and flow phantoms.