The HML's new field deployable, high-resolution whole body counter
Gary H Kramer1, Kevin Capello, Barry M Hauck
1Human Monitoring Laboratory, Radiation Surveillance and Heath Assessment Division, Radiation Protection Bureau, 775 Brookfield Road, PL6302D1, Ottawa, Ontario K1A 1C1, Canada.
The Ortec Detective instrument, repurposed as a portable whole body counter, shows adequate sensitivity for detecting fission and activation products during emergencies. Calibration using Monte Carlo simulations ensures accurate measurements for different body sizes.
Area of Science:
- Nuclear instrumentation
- Radiation detection and measurement
- Health physics
Background:
- The Human Monitoring Laboratory (HML) sought alternative applications for a hyperpure germanium field instrument, the Ortec Detective.
- Originally designed for detecting radioactive contraband, its potential as a portable whole body counter was explored.
Purpose of the Study:
- To evaluate the Ortec Detective's suitability as a high-resolution portable whole body counter for emergency response.
- To determine the instrument's sensitivity for various radioactive isotopes, including fission products, activation products, and actinides.
- To develop calibration methods for accurate measurements based on photon energy and subject size.
Main Methods:
- Repurposing the Ortec Detective instrument by connecting it to a laptop to function as a spectroscopy system.
- Utilizing Monte Carlo simulations for instrument calibration and optimization of counting geometry.
- Developing empirical functions to relate counting efficiency to photon energy, subject size, and height.
Main Results:
- The instrument demonstrated adequate sensitivity for detecting fission and activation products but not actinides in emergency scenarios.
- Calibration curves were established as a function of photon energy and subject size for whole body counting.
- A function correlating counting efficiency with subject height was developed for thyroid counting.
- Minimum Detectable Activities (MDAs) were found to be in the low kilo Becquerel range for fission and activation products with a 5-minute count.
Conclusions:
- The Ortec Detective, when configured as a portable whole body counter, is a viable tool for emergency response concerning fission and activation products.
- Monte Carlo simulations and empirical functions provide effective calibration strategies for this repurposed instrument.
- Further investigation is needed to assess its utility for actinide detection.