Comparative evaluation of an II based and a flat panel based cardiovascular fluoroscopy system within a clinical
1Medical Physics Department, Westmead Hospital, Westmead, Australia. rrgr@imag.wsahs.nsw.gov.au
Abstract:
The image quality and dose parameters from a 2004 Siemens Axiom Artis dBC cardiac biplane with flat panel detector were evaluated and compared to similar parameters evaluated for a 1977 Toshiba DPF 2000A biplane cardiac unit with a conventional image intensifier. Image quality assessment was performed with the Westmead test object; using solid water as a patient equivalent absorber. The patient dose comparison of the two systems is based on dose area product meter readings for 1512 patient cases recorded over 6 months following installation of the Siemens flat panel digital unit. The image quality results indicate that: (a) high contrast resolution was better with the digital flat panel unit, (b) low contrast resolution is similar between systems, and (c) the threshold contrast of the flat panel system is the same or inferior to that of the image intensifier system. Input dose to the surface of the flat panel detector showed a strong dependence on field size, similar to the behaviour of image intensifier system. For the most common clinical procedure--Left Heart Study via Judkins--the average total dose area product reading was 64.0 Gy-cm2 against 67.7 Gy-cm2 for the digital and conventional units respectively (p = 0.27) indicating no significant difference in dose performance between the two x-ray machines.
More Related Videos
07:43In Vivo Quantification of Hip Arthrokinematics during Dynamic Weight-bearing Activities using Dual Fluoroscopy
Published on: July 2, 2021
03:47Workflow and Framework for Collecting and Implementing Point-of-Care Ultrasound Data in the Management of Heart Failure Patients
Published on: July 12, 2024
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System V: CT