Measurement of CT scanner dose profiles in a filmless department
1Radiology Department, Auckland City Hospital, Auckland, New Zealand. fthomson@adhb.govt.nz
Abstract:
The measurement of the FWHM of the slice thickness radiation dose profile of a CT scanner using a prototype low sensitivity CR imaging plate has been investigated, as an alternative to the traditional method using envelope-packed industrial film. Using a standard Agfa clinical CR system to acquire the image, the FWHM of the dose profile can be accurately measured using readily available Public Domain software. An Agfa 18 x 24 cm CR cassette gives a pixel pitch of 113.5 microm, but with interpolation, the measurement accuracy can be less than 1 pixel. For a nominal 10 mm collimation, 15 successive measurements of the FWHM using CR gave an average width of 10.00 mm with a standard deviation of 0.02 mm. This may be compared with 4 successive measurements using film and a dual exposure technique to define the optical density at half peak height, yielding an average width of 9.98 mm with a SD of 0.03 mm. This prototype NDT plate is not a commercial product, but a radiotherapy plate with a similar sensitivity is available commercially and should give similar results.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
X-ray Imaging