Full-root Aortic Valve Replacement by Stentless Aortic Xenografts in Patients with Small Aortic Roots
Published on: May 21, 2017
The On-X heart valve: mid-term results in a poorly anticoagulated population
Mervyn A Williams1, Sonia van Riet
1Provincial Hospital, Port Elizabeth, South Africa. drmwilliams@telkomsa.net
Background And Aim Of The Study:
The study aim was to evaluate the clinical performance of the On-X heart valve in a socioeconomically disadvantaged population. Most patients were from an indigenous, poorly educated and geographically dispersed segment of the population where anticoagulation therapy was generally erratic.
Methods:
Between 1999 and 2004, a total of 530 valves (242 mitral valves, 104 aortic valves, 92 double valves) was implanted in 438 patients (average age 33 years; range: 3-78 years). The most common reason for surgery was rheumatic valve disease (57%), followed by degenerative valve disease (11%) and infective endocarditis (9%). Follow up was 95% complete for a total of 746 patient-years (pt-yr). Among the patient population, 40% were either not anticoagulated or were unsatisfactorily anticoagulated.
Results:
Hospital mortality was 2.3%, and none of the hospital deaths was valve-related. Mean (+/- SE) actuarial survival (including hospital deaths) at four years was: AVR 73.8 +/- 8.1%, MVR 83.4 +/- 5.7% and DVR 60.9 +/- 10.3%. Linearized rates (for AVR, MVR and DVR, respectively) for late complications (%/ pt-yr) were: bleeding events 0.6, 1.0, and 2.3; thrombosis 0.0, 0.2, and 0.0; endocarditis 0.6, 1.0, and 2.3; paravalvular leak 0.6, 0.2, and 0.0; systemic embolism 1.1, 1.5, and 3.5. Most systemic emboli were related to infective endocarditis. Among patients there were seven uncomplicated, full-term pregnancies.
Conclusion:
Bearing in mind the erratic anticoagulation coverage and high incidence of infective endocarditis, the results of this study may be regarded as encouraging. The low incidence of valve thrombosis (one case) was noteworthy. These data also suggest that the On-X valve may be implanted with relative safety in women wishing to have children.
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Mitral Stenosis I: Introduction
Mitral Stenosis II: Clinical features and Diagnostic Tests
Anticoagulant Drugs: Low-Molecular-Weight Heparins
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Mitral Stenosis IV: Nursing Management