Sensitive quantitative analysis of C-peptide in human plasma by 2-dimensional liquid chromatography-mass spectrometry
Eduard Rogatsky1, Beate Balent, Gayotri Goswami
1Department of Medicine and General Clinical Research Center at Albert Einstein College of Medicine of Yeshiva University, Bronx, NY 10461, USA.
Background:
Isotope-dilution assays (IDAs) are well established for quantification of metabolites or small drug molecules in biological fluids. Because of their increased specificity, IDAs are an alternative to immunoassays for measuring C-peptide.
Methods:
We evaluated a 2-dimensional liquid chromatography-mass spectrometry (2D LC/MS) IDA method. Sample preparation was by off-line solid-phase extraction, and C-peptide separation was performed on an Agilent 1100 2D LC system with a purification method based on high-pressure switching between 2 high-resolution reversed-phase columns. Because of the low fragmentation efficiency of C-peptide, multiple-reaction monitoring analysis was omitted and selective-ion monitoring mode was chosen for quantification. Native and isotope-labeled ([M+18] and [M+30]) C-peptides were monitored in the +3 state at m/z 1007.7, 1013.7, and 1017.7.
Results:
The assay was linear (r(2) = 0.9995), with a detection limit of 300 amole (1 pg) on column. Inter- and intraday CVs for C-peptide were < or =2%. Comparison with an established polyclonal-based RIA showed high correlation (r = 0.964). Plasma concentrations of total C-peptide measured by RIA were consistently higher than by IDA LC/MS, consistent with the higher specificity of IDAs compared with immunoassays.
Conclusions:
The 2D LC/MS IDA approach eliminates matrix effects, enhancing assay performance and reliability, and has a detection limit 100-fold lower than any previously reported LC/MS method. Isotope-labeled C-peptide(s) can be clearly differentiated from endogenous C-peptide by the difference in m/z ratio, so that both peptides can be quantified simultaneously. The method is highly precise, robust, and applicable to pharmacokinetic detection of plasma peptides.