External quality assessment scheme in red blood cell serology: a 5-year experience in Thailand
S Bejrachandra1, J Saipin, O Nathalang
1Department of Transfusion Medicine, Faculty of Medicine, Siriraj Hospital, Mahidol University, Bangkok 10700, Thailand.
From 2000 to 2004, 36, 58, 72, 78, and 86 laboratories participated in an external quality assessment scheme (EQAS) organized by the Department of Transfusion Medicine, Faculty of Medicine Siriraj Hospital. Each year the staff was requested to perform ABO grouping, D typing, antibody screening, antibody identification, and DATs on eight blood samples. Each participant received information on the correct test results and a coded summary. Regarding ABO grouping, the error rate ranged from 0.3 to 1.3 percent, mostly due to human errors. Error rates in D typing ranged from 0.7 to 5.7 percent, the most problematic being weak D phenotype interpretation. Although every sample was negative by the DAT, error rates due to false positive test results were determined to be 0.4 to 2.1 percent. Antibody screening errors were also found; however, errors steadily decreased from 4.2 percent in 2000 to 0.3 percent in 2004. Only 69.4 to 87.2 percent of laboratories performed antibody identification; however, correct results increased from 78.4 to 91.0 percent. In conclusion, an EQAS in RBC serology should be used to compare results from different laboratories and to identify those laboratories that need improvement in testing procedures.
From 2000 to 2004, 36, 58, 72, 78, and 86 laboratories participated in an external quality assessment scheme (EQAS) organized by the Department of Transfusion Medicine, Faculty of Medicine Siriraj Hospital. Each year the staff was requested to perform ABO grouping, D typing, antibody screening, antibody identification, and DATs on eight blood samples. Each participant received information on the correct test results and a coded summary. Regarding ABO grouping, the error rate ranged from 0.3 to 1.3 percent, mostly due to human errors. Error rates in D typing ranged from 0.7 to 5.7 percent, the most problematic being weak D phenotype interpretation. Although every sample was negative by the DAT, error rates due to false positive test results were determined to be 0.4 to 2.1 percent. Antibody screening errors were also found; however, errors steadily decreased from 4.2 percent in 2000 to 0.3 percent in 2004. Only 69.4 to 87.2 percent of laboratories performed antibody identification; however, correct results increased from 78.4 to 91.0 percent. In conclusion, an EQAS in RBC serology should be used to compare results from different laboratories and to identify those laboratories that need improvement in testing procedures.
More Related Videos
10:50Detection and Quantification of Plasmodium falciparum in Aqueous Red Blood Cells by Attenuated Total Reflection Infrared Spectroscopy and Multivariate Data Analysis
Published on: November 2, 2018
11:27Prediction of Red Blood Cell Antibody Significance Using the Monocyte-Macrophage Assay
Published on: February 7, 2025