Related Experiment Video
Updated: Jul 25, 2026

Angle-resolved Photoemission Spectroscopy At Ultra-low Temperatures
Published on: October 9, 2012
Performance limits of direct cryogenically cooled silicon monochromators - experimental results at the APS
W K Lee1, P Fernandez, D M Mills
1Advanced Photon Source, Argonne National Laboratory, 9700 South Cass Avenue, Argonne, IL 60439, USA. wklee@aps.anl.gov
Abstract:
The successful use of cryogenically cooled silicon monochromators at third-generation synchrotron facilities is well documented. At the Advanced Photon Source (APS) it has been shown that, at 100 mA operation with the standard APS undulator A, the cryogenically cooled silicon monochromator performs very well with minimal (<2 arcsec) or no observable thermal distortions. However, to date there has not been any systematic experimental study on the performance limits of this approach. This paper presents experimental results on the performance limits of these directly cooled crystals. The results show that if the beam is limited to the size of the radiation central cone then, at the APS, the crystal will still perform well at twice the present 100 mA single 2.4 m-long 3.3 cm-period undulator heat load. However, the performance would degrade rapidly if a much larger incident white-beam size is utilized.

