Related Experiment Video
Updated: Aug 17, 2026

An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016
Polyacetal rod fixation of fractures in osteoporotic bone. A preliminary report
1Royal North Shore Hospital, Sydney, St Leonards, New South Wales, Australia.
Abstract:
Intramedullary rods manufactured from polyacetal were used to fix diaphyseal fractures in osteoporotic bone. They are pliable and can be introduced without further damage to such bone. Their low elastic modulus induces abundant callus. They can be locked with cortical bone screws without the need of jigs or radiographic control. Nine femoral and five tibial fractures were fixed in 10 patients, all with osteoporosis; 13 united primarily; one required bone grafting. No implant broke; however, no ambulant patient weighed more than 75 kg. The rods offer a promising treatment for diaphyseal fractures in osteoporotic bone.