Thermal stability of glucose and other sugar aldoses in normal phase high performance liquid chromatography
Rune Slimestad1, Ingunn M Vågen
1PlantChem, Saerheim Research Centre, N-4353 Klepp station, Norway.
Abstract:
Analysis of glucose and other simple sugars are often performed by use of normal phase HPLC methods with acetonitrile as major eluent. The present results clearly show that column temperature plays an important role with respect to chromatographic performance and detection limits of glucose when using a specific carbohydrate column. A change in column temperature from 25 to 45 degrees C reduced the detection of glucose (with ELSD) by more than 41%, whereas the detection of other sugar aldoses (galactose, xylose and rhamnose) were suppressed even more. By increase of column temperature to 70 degrees C the detector signal of glucose was found to be less than 2% compared to that obtained at 20 degrees C. Neither fructose nor sucrose showed similar correlation between column temperature and detection. The rate of decreased response is not dependent on sample concentration or the ELSD settings. The results express the importance of accurate temperature control in the analysis of sugar aldoses, and also values low column temperatures for samples with low concentrations of sugar aldoses in order to improve detection.
Related Concept Videos
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...
High-Performance Liquid Chromatography: Instrumentation
Silica Gel Column Chromatography: Overview
Polar components tend to bind strongly to the silica gel, causing them to move slowly through the column. In contrast, nonpolar compounds...