Passive catheter visualization in magnetic resonance-guided endovascular therapy using multicycle projection
Jonathan N Draper1, M Louis Lauzon, Richard Frayne
1Department of Electrical and Computer Engineering, University of Calgary, Calgary, Alberta, Canada.
Purpose:
To improve upon the conventional projection dephaser (PD) method of background suppression and evaluate the use of multicycle projection dephasers to improve catheter conspicuity in background-suppressed MR images.
Materials And Methods:
Passive visualization of endovascular catheters in MR images is compared using two background suppression techniques: 1) the conventional PD method and 2) the multicycle PD method. Contrast-filled 4-French (1.3 mm) catheters were imaged in homogeneous and heterogeneous phantoms, and in the common carotid artery of a canine using a modified spoiled gradient echo imaging sequence. We used catheter-to-background contrast (ranging from -100% to 100%) as the metric to compare background suppression techniques.
Results:
In the homogeneous and heterogeneous phantoms, the contrast was -6.9% (catheter darker than background) and 15.0%, respectively, using the conventional PD method, and 50.6% and 44.0%, respectively, using the multicycle PD method. In the canine carotid artery, the contrast was -3.1% using the conventional PD method and 53.0% using the multicycle PD method.
Conclusion:
This work shows that multicycle projection dephasers improve catheter conspicuity over the conventional PD method. The multicycle PD method has potential for use in guiding endovascular procedures.
Insights
Multicycle projection dephasers significantly enhance catheter visibility in MRI scans compared to conventional methods. This advancement improves the conspicuity of endovascular catheters, aiding in procedures.
Area of Science:
- Medical Imaging
- Biomedical Engineering
- Interventional Radiology
Background:
- Conventional projection dephaser (PD) methods have limitations in background suppression for MR imaging.
- Improving catheter conspicuity is crucial for guiding endovascular procedures.
Purpose of the Study:
- To evaluate multicycle projection dephasers for improved background suppression.
- To enhance the conspicuity of endovascular catheters in MR images.
Main Methods:
- Comparison of conventional PD and multicycle PD techniques.
- Imaging of catheters in phantoms and canine carotid arteries.
- Utilizing a modified spoiled gradient echo imaging sequence.
Main Results:
- Multicycle PD yielded significantly higher catheter-to-background contrast (50.6% in homogeneous, 44.0% in heterogeneous phantoms) than conventional PD (-6.9% and 15.0%).
- In vivo, multicycle PD achieved 53.0% contrast versus -3.1% for conventional PD in canine carotid arteries.
Conclusions:
- Multicycle projection dephasers offer superior catheter conspicuity compared to conventional PD.
- This technique shows promise for enhancing guidance in endovascular procedures.