Related Experiment Video
Updated: Aug 7, 2026

The Portable Chemical Sterilizer (PCS), D-FENS, and D-FEND ALL: Novel Chlorine Dioxide Decontamination Technologies for the Military
Published on: June 29, 2014
Redesign of portable suction equipment cases: an engineering approach to a disinfection problem
Alice N Neely1, Lisa A Regnold, Susan Hayes
1Department of Infection Control, University of Cincinnati College of Medicine, USA. aneely@shrinenet.org
Background:
Patients with tracheostomies carry a case for suctioning supplies. A returned case cultured positive for Staphylococcus aureus even after it had been cleaned with a disinfectant. The purposes of this study were to determine (1) the source of the S aureus and (2) how to provide a microbiologically safe case for the next patient.
Methods:
Reviews of patient and environmental cultures plus pulsed-field gel electrophoresis (PFGE) were used to determine the source of the S aureus. Repeated wiping with various detergents/disinfectants did not remove the S aureus from the case, which could not be immersed for cleaning because of a chipboard component. We designed a case made of denim with a washable synthetic fiber to provide shape. The new cases were purposefully contaminated and laundered and then cultured.
Results:
PFGE indicated that the S aureus was from the patient who had most recently used the case. Contamination of the denim case with the original S aureus, followed by laundering, resulted in a case that was free of the S aureus. These results were repeated for other pathogens.
Conclusion:
Commercially available cases for suction equipment can become contaminated with pathogens from the user and may be difficult to disinfect. By making the case out of washable materials, less expensive cases, which could be readily disinfected, resulted.
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Handwashing I: Introduction and Types of Equipment
Hand wash basins in clinical areas should have faucets that can be turned on and off without using the hands; that is, they should be non-touch or lever-operated.
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...