Increasing left ventricular pacing output decreases interventricular conduction time in patients with biventricular
William H Sauer1, Jonathan S Sussman, Ralph J Verdino
1Electrophysiology Section, Division of Cardiovascular Medicine, University of Pennsylvania Health System, Philadelphia, Pennsylvania, USA. william.sauer@uphs.upenn.edu
Objective:
To evaluate the effect of increasing LV pacing output on interventricular timing in patients with biventricular pacing systems.
Background:
Clinical improvement with biventricular pacing is likely related to reduction in ventricular dysynchrony in patients with cardiomyopathy. We hypothesized that increasing left ventricular pacing output would reduce interventricular conduction time and could affect ventricular synchrony.
Methods:
Forty-two sequential patients with biventricular pacing systems that permitted independent LV pacing were selected at the time of routine device interrogation. The interval between LV pacing stimulus and onset of the RV electrogram was measured during LV pacing at capture threshold and at maximum pacing output for each patient.
Results:
The average time from LV pacing stimulus to right ventricular electrogram onset was 142.5 +/- 32.5 ms (range 90-230 ms) at threshold and 132.3 +/- 30.4 ms (range 90-220 ms) at maximum pacing output, with a mean decrease in conduction time of 10.2 +/- 10.9 ms (range 0-45 ms). There was significantly greater interventricular conduction shortening with increased pacing output in patients with ischemic cardiomyopathy compared to others (14.9 +/- 11.9 ms vs 4.0 +/- 4.6 ms; P < 0.01).
Conclusions:
Conduction time from LV to RV shortens as LV pacing output is increased. This effect is seen to a greater degree in patients with ischemic cardiomyopathy, possibly related to the presence of myocardial scar near the pacing electrode. Further investigation is needed to assess the clinical outcomes related to this new method for optimizing resynchronization therapy.
More Related Videos
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Mitral Stenosis I: Introduction
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Conduction System of the Heart
This system relies on the unique properties of nodal and Purkinje cells:...
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send blood...
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...