Gastrointestinal endoscopes cleaned without detergent substance following an automated endoscope washer/disinfector
Marie Méan1, Marie Reine Mallaret, Philippe Bichard
1Unité d'hygiène hospitalière, Centre Hospitalier Universitaire de Grenoble, Grenoble.
Objective:
To report cases of gastrointestinal endoscopies performed with endoscopes that were reprocessed without detergent substance during a period of dysfunction of the automated endoscope reprocessor (AER).
Method:
A dysfunction of the AER for the cycles requiring detergent substance was reported at the Grenoble University Hospital on March 2005. During this period, 72 patients had potentially been exposed to a contaminated endoscope. A recall procedure was organized and serologic tests (HIV, HCV, HBV) were performed 3 and 6 months after the AER incident.
Results:
Within the 72 patients convened, 56 (77.8%) were seen in consultation and accepted the serologic screening. Finally, serologic screening was done for 59 patients (81.9%) and no seroconversion for HIV, HCV, or HBV was observed. The final attrition rate was 13 patients (18.1%).
Conclusion:
No viral infection was transmitted during the AER dysfunction. After this AER incident, the monitoring of the endoscopic procedures and the traceability of the cleaning process were both improved to prevent further incidents.
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Handwashing I: Introduction and Types of Equipment
Hand wash basins in clinical areas should have faucets that can be turned on and off without using the hands; that is, they should be non-touch or lever-operated.
Hand hygiene
Hand washing...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Handwashing III: During the Procedure and Post-Procedure Steps