CW-OSL measurement protocols using optical fibre Al2O3:C dosemeters
J M Edmund1, C E Andersen, C J Marckmann
1Risø National Laboratory, Radiation Research Department, DK-4000 Roskilde, Denmark. jens.edmund@risoe.dk
Radiation Protection Dosimetry
|September 23, 2006
Summary
A new in vivo dosimetry system uses carbon-doped aluminium oxide (Al2O3:C) crystals and continuous wave optically stimulated luminescence (CW-OSL) for accurate radiotherapy dose measurement. The protocol enhances reproducibility and linearity by analyzing specific luminescence decay components.
Area of Science:
- Medical Physics
- Radiation Oncology
- Dosimetry
Background:
- Accurate in vivo dosimetry is crucial for effective radiotherapy.
- Traditional dosimetry methods face challenges in real-time dose monitoring.
- Carbon-doped aluminium oxide (Al2O3:C) offers potential for luminescence-based dosimetry.
Purpose of the Study:
- To introduce and validate a new system for in vivo dosimetry during radiotherapy.
- To establish a reproducible and linear measurement protocol for clinical application.
- To optimize the analysis of luminescence signals for precise dose determination.
Main Methods:
- Utilized a novel system employing Al2O3:C crystals and optical fibers for signal transmission.
- Employed continuous wave optically stimulated luminescence (CW-OSL) with a 532 nm laser.
- Developed a measurement protocol involving signal stabilization with a minimum pre-dose and analysis of luminescence decay components.
Main Results:
- Demonstrated high reproducibility and improved linearity of the CW-OSL measurement protocol.
- Identified the necessity of a crystal-specific minimum pre-dose for signal stabilization.
- Showcased the effectiveness of separating decay curves into components, utilizing only fast and medium components for accurate measurement.
Conclusions:
- The developed CW-OSL system and protocol are suitable for clinical in vivo dosimetry.
- Optimized signal processing, including component analysis of decay curves, is essential for accurate dose measurement.
- This advancement offers a promising tool for real-time radiotherapy dose monitoring and quality assurance.