Lorena Bouzas1, J Carlos Tutor
1Unidad Monitorización Fármacos, Laboratorio Central, Hospital Clínico Universitario, Santiago de Compostela, Spain.
Objectives:
Everolimus (Certican) is a new immunosuppressant derived from sirolimus (Rapamune) with a 2-hydroxyethyl chain at position 40 of the macrolide ring. The aim of our study was to evaluate the possible determination of everolimus in whole blood using a commercialized microparticle enzyme immunoassay for sirolimus determination.
Design And Methods:
Everolimus concentrations were determined in blood samples from 11 kidney transplant patients (n=51) and different control materials (n=35) using the Seradyn Innofluor Certican fluorescence polarization immunoassay (FPIA) and the Abbott IMx sirolimus microparticle enzyme immunoassay (MEIA).
Results:
The MEIA gave a concentration-dependent cross-reactivity with the everolimus, with a linear regression between the assigned values (y) for the Innofluor Certican calibrators and those obtained (x) using this immunoassay: y=0.96x+0.67 (ma68=0.21 microg/L, r=0.974, p<0.001). The within- and between-run coefficients of variation using the MEIA were
Conclusions:
The Abbott IMx sirolimus MEIA permits the simple and precise determination of everolimus in whole blood, making it a valid alternative for the therapeutic monitoring of this immunosuppressant agent.
More Related Videos
07:52Detection of Antibodies That Neutralize the Cellular Uptake of Enzyme Replacement Therapies with a Cell-based Assay
Published on: September 10, 2018
08:26Development and Validation of an Ultrasensitive Single Molecule Array Digital Enzyme-linked Immunosorbent Assay for Human Interferon-α
Published on: June 14, 2018
Related Concept Videos
Therapeutic Drug Monitoring: Drug Analysis Methods
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations