The aim of this study was to assess the accuracy of cardiac functional values obtained from free-breathing real-time cine CMR with the temporal sensitivity encoding (TSENSE) technique by comparing them with values obtained from conventional cine CMR. For the real-time cine CMR, two protocols were employed, one with good temporal resolution and one with good spatial resolution. The functional values obtained from the high temporal resolution real-time cine CMR agreed and correlated well with those of cine CMR. On the other hand, statistically significant but clinically slight overestimation of ESV (p < .05) and underestimation of EF (p < .01) were observed with the other protocol. Real-time cine CMR with TSENSE can provide acceptable cardiac functional values.