Measurement of cardiac output using Physio Flow with different positions of electrode placement
1Department of Anaesthesia and Surgical Intensive Care, Singapore General Hospital, Outram Road, Singapore. tankianhian@gmail.com
Introduction:
Physio Flow is a non-invasive impedance cardiograph device that measures cardiac output. Recommended electrode placements involve six electrodes, including two near the xiphisternum (Z3 and Z4/ ECG3/neutral). This study aims to evaluate if changing the positions of these two leads to the left fourth and fifth intercostal spaces along the mid-axillary line results in a change in the cardiac output measurement.
Methods:
This was a prospective, controlled, crossover, paired study of 30 patients where electrodes were placed in the recommended positions and cardiac output (CO1) obtained after two minutes. The second cardiac output (CO2) was then obtained with the electrodes Z3 and Z4/ECG3/neutral repositioned at the left mid-axillary line at the fourth and fifth intercostal spaces. The final step involved switching the Z3 and Z4/ECG3/neutral leads back to the recommended position and the cardiac output (CO3) was measured.
Results:
The average of the initial and third readings (COave) was compared with the measured CO2 and analysed. The regression equation was: CO at the proposed site (CO2) = COave at the recommended site + 0.058. The paired samples correlation was 0.995. Within the 95 percent limits of agreement, the bias with CO measured at the proposed site of electrode placement was 0.046 L/min with the limits at -0.24 L/min and 0.34 L/min. The mean difference was 0.86% of the average CO.
Conclusion:
A small positive bias was demonstrated when Physio Flow measurements were taken with the leads Z3 and Z4/ECG3/neutral placed in the mid-axillary line fourth and fifth intercostal spaces.
Related Concept Videos
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls between...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...