Related Experiment Video
Updated: Jul 18, 2026

Applying Dynamic Strain on Thin Oxide Films Immobilized on a Pseudoelastic Nickel-Titanium Alloy
Published on: July 28, 2020
Effect of human organism on the oxide layer formed on titanium osteosynthesis plates: a surface analytical study
Norbert Velich1, Bence Kádár, Gábor Kiss
1Semmelweis University, Faculty of Dentistry, Department of Oral and Maxillofacial Surgery and Dentistry, Budapest, Hungary.
Abstract:
The question of whether or not to remove the titanium osteosynthesis plates used in maxillofacial surgery is not yet answered. These plates can be left a long time in the organism if there is no harmful interaction between the plates and the organism. The authors examined the properties of the surface oxide layers formed on osteosynthesis titanium plates by subsequent thermal and anodic oxidation, together with the properties of plates removed from patients after three years. Surface analytical method Secondary Ion Mass Spectroscopy (SIMS) was chosen. It was found that inside of the about 200-mum thick oxide on the original plates the concentration of impurities is much lower than near the oxide/titanium interface. On the surface of plates removed from the human body, a C (carbon), Ca (calcium) and P (phosphorus) enrichment was detected, suggesting a biological interaction between the organism and the plate. The passivating layer formed with thermal and anodic oxidation has not changed significantly in three years; it resisted to the corrosive effect of the human organism.