Related Experiment Video
Updated: Jul 18, 2026

Assessing Early Stage Open-Angle Glaucoma in Patients by Isolated-Check Visual Evoked Potential
Published on: May 25, 2020
Glaucoma detection with the Heidelberg retina tomograph 3
Zvia Burgansky-Eliash1, Gadi Wollstein, Richard A Bilonick
1UPMC Eye Center, Eye and Ear Institute, Ophthalmology and Visual Science Research Center, Department of Ophthalmology, University of Pittsburgh School of Medicine, Pittsburgh, Pennsylvania 15213, USA.
The Heidelberg Retina Tomograph 3 (HRT 3) software shows similar glaucoma detection capabilities to HRT 2. New HRT 3 glaucoma probability score analysis effectively distinguishes healthy from glaucomatous eyes without subjective input.
Area of Science:
- Ophthalmology
- Medical Imaging
- Glaucoma Research
Background:
- Glaucoma diagnosis relies on detecting optic nerve damage.
- Heidelberg Retina Tomograph (HRT) devices aid in quantifying optic nerve head parameters.
- Comparing HRT software versions is crucial for diagnostic accuracy.
Purpose of the Study:
- To evaluate and compare the diagnostic performance of Heidelberg Retina Tomograph version 3 (HRT 3) software against HRT version 2 (HRT 2).
- To assess the ability of both HRT 2 and HRT 3 to differentiate between healthy and glaucomatous eyes.
Main Methods:
- A retrospective cross-sectional study involving 71 healthy eyes and 50 glaucomatous eyes.
- HRT 2 data were analyzed using HRT 3 software.
- Discrimination was assessed using areas under the receiver operating characteristics (AROCs) curves.
Main Results:
- Individual stereometric parameters showed good discrimination, with linear cup/disc ratio and horizontal retinal nerve fiber layer curvature performing well.
- Overall classification by HRT 2 Moorfields regression analysis (MRA), HRT 3 MRA, and HRT 3 glaucoma probability score (GPS) yielded comparable AROCs (0.927, 0.934, and 0.880, respectively).
- Agreement between HRT 2 and HRT 3 MRA was good (kappa = 0.70), with HRT 3 identifying more abnormalities.
Conclusions:
- The glaucoma discriminating ability of HRT 3 software is comparable to HRT 2.
- HRT 3 glaucoma probability score analysis demonstrates potential for objective differentiation between healthy and glaucomatous eyes.
Related Concept Videos
Glaucoma: Overview
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...