In vitro evaluation of forces exerted by a new computer-assisted colonoscope (the NeoGuide Endoscopy System)
A Eickhoff1, R Jakobs, A Kamal
1Medical Department C, Klinikum Ludwigshafen, Academic Hospital of the University of Mainz, Ludwigshafen, Germany. eickhofa@klilu.de
Background:
The NeoGuide Endoscopy System (NES) utilizes a fully articulated, computer-controlled insertion tube that allows proximal segments of the colonoscope to follow the path taken by the tip as it is manually advanced through the colon. The system was designed to eliminate looping and scope displacement during colonoscopy.
Material And Methods:
Using in vitro testing, an inanimate flexible model of the colon incorporating four force transducers located at the key flexure points was employed to measure the axial forces on the colon wall during colonoscopy. In the second part of the study, 10 gastroenterologists performed colonoscopies, using a training latex-based simulator, with the NES and with a standard colonoscope. Colonic displacement was independently assessed by six gastroenterologists, with each evaluating endoscopist assigning a score between 0 and 5 corresponding to the maximum colonic displacement observed at any location.
Results:
The average measured forces (in lbs) at three of the four flexure points were significantly lower ( P < 0.05) when the NES was used. The mean colonic displacement was significantly lower for procedures performed with the NES compared with the standard colonoscope (2.36 vs. 4.26, P < 0.001). Interobserver agreement regarding the degree of colonic displacement due to looping was moderate (weighted kappa = 0.45, P < 0.01).
Conclusions:
Colonoscopy with the NES was associated with significantly less looping and lateral force required for advancement than procedures with a standard colonoscope. The reduced amount of looping suggests that use of the NES in patients might be associated with less discomfort and thus require less sedation.
Related Concept Videos
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and solid...
Endoscopic Procedures II: Colonoscopy
Endoscopic Procedures III: Video Capsule Endoscopy
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to: