Related Experiment Video
Updated: Jul 18, 2026

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
High-performance liquid chromatography-electrospray ionization mass spectrometry determination of mitiglinide in
Jiabi Liang1, Yuan Tian, Zunjian Zhang
1Key Laboratory of Drug Quality Control and Pharmacovigilance (China Pharmaceutical University), Ministry of Education, Nanjing, 210009, China.
Abstract:
A selective and sensitive method employing high-performance liquid chromatography-electrospray ionization mass spectrometry was developed and validated for the determination of mitiglinide in human plasma.With gliclazide as the internal standard, mitiglinide was extracted from plasma with n-hexane: = 80 : 20 (v/v). The organic layer was evaporated and the residue was redissolved in methanol: water (10 mM CH3COONH4, pH = 3.0) = 65 : 35 (v/v). An aliquot of 10 microl was chromatographically analyzed on a prepacked Shimadzu VP-ODS (5 microm, 150 x 2.0 mm i.d.) using the mobile phase comprising methanol: water (10 mM CH3COONH4) = 65 : 35 (v/v) by means of selected-ion monitoring mode mass spectrometry. Standard curves were linear (r2 = 0.9972) over the concentration range of 2.84-11 300 pmol/ml and had good accuracy and precision. The within- and between-batch precisions of the method were within 15% of standard deviation. The lower limit of detection was 1.42 pmol/ml. The validated HPLC/ESI-MS method has been successfully applied in the pharmacokinetics of mitiglinide in 12 healthy Chinese volunteers.