Unusual pacing observed in a biventricular pacemaker with epicardial pacing leads
William H Sauer1, Fermin C Garcia, Ralph J Verdino
1University of Pennsylvania Health System, Philadelphia, Pennsylvania, USA.
Abstract:
In this case report, we describe inappropriate pacing in a patient with a newly implanted biventricular pacing system with three epicardial leads. Very high pacing thresholds resulted from the use of nonsteroid eluting leads, and unusual pacing behavior was due to the chosen implant locations for the right atrial and right ventricular epicardial leads. The interplay between the programmed pacemaker parameters, the very late ventricular sensing in the setting of intrinsic right bundle branch block, and atrial lead oversensing led to ineffective biventricular pacing. The persistently elevated pacing thresholds and sensing problems led to very early battery depletion and the need to replace the epicardial system with three new endocardial leads.
Related Concept Videos
Dysrhythmias III: Characteristics of Dysrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Electrophysiology of Normal Cardiac Rhythm
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...