Impact of a pressure-responsive flow-limiting valve on bag-valve-mask ventilation in an airway model
Jonnathan M Busko1, Thomas H Blackwell
1Department of Emergency Medicine, Carolinas Medical Center, Charlotte, North Carolina 28203, USA. jbuskomd@yahoo.com
Objective:
Using a simulated airway model, we compared ventilation performance by emergency medical services (EMS) providers using a traditional bag-valve-mask (Easy Grip) resuscitator to their performance when using a new device, the SMART BAG resuscitator, which has a pressure-responsive flow-limiting valve.
Methods:
We recruited EMS providers at an EMS educational forum and performed a randomized, non-blinded, prospective crossover comparison of ventilation with 2 devices on a non-intubated simulated airway model. Subjects were instructed to ventilate a Mini Ventilation Training Analyzer as they would an 85-kg adult patient in respiratory arrest. After being randomized to order of device use, they performed ventilation for 1 minute with each device. Primary outcomes were ventilation rates and peak airway pressures. We also measured average tidal volume, gastric inflation volume, minute ventilation and inspiratory:expiratory (I:E) ratio, and compared our results to the American Heart Association standards (2005 edition).
Results:
We observed statistically significant differences between the SMART BAG and the traditional bag-valve-mask for respiratory rate (12 v. 14 breaths/min), peak airway pressure (15.6 v. 18.9 cm H(2)O), gastric inflation (239.6 v. 1598.4 mL), minute ventilation (7980 v. 8775 mL), and I:E ratio (1.3 v. 1.1). Average tidal volume was similar with both devices (679.6 v. 672.2 mL).
Conclusion:
The SMART BAG(R) provided ventilation performance that was more consistent with American Heart Association guidelines and delivered similar tidal volumes when compared with ventilation with a traditional bag-valve-mask resuscitator.
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)