Related Experiment Video
Updated: Jul 6, 2026

Noninvasive In Vivo Small Animal MRI and MRS: Basic Experimental Procedures
Published on: October 20, 2009
Inversion recovery prepared turbo spin echo sequences with reduced SAR using smooth transitions between pseudo steady
Matthias Weigel1, Maxim Zaitsev, Juergen Hennig
1Department of Diagnostic Radiology, Medical Physics, University Hospital Freiburg, Freiburg, Germany. Matthias.Weigel@uniklinik-freiburg.de
Abstract:
This study investigates the contrast behavior of 2D inversion recovery (IR) prepared turbo spin echo (TSE) sequences that use RF pulse schemes with variable low flip angles (hyperTSE) to reduce RF power deposition. A framework of equations and calculations for adapting the sequence parameters is presented by which equivalent image contrast is achieved compared to conventional IR-TSE imaging. Although the inversion time (TI) and repetition time (TR) do not need to be changed, the echo time (TE) has to be prolonged such that the effective TE (TE(eff)) is preserved. Measurements in healthy volunteers confirmed this finding for IR-TSE sequences using different TIs: fluid attenuated inversion recovery (FLAIR), gray matter (GM)-white matter (WM)-IR, and short tau IR (STIR). The results demonstrate that hyperTSE sequences enable high-quality IR-prepared imaging with a considerably reduced specific absorption rate (SAR).
Related Concept Videos
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are slanted or...
NMR Spectrometers: Resolution and Error Correction
NMR Spectroscopy: Spin–Spin Coupling
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Insensitive Nuclei Enhanced by Polarization Transfer (INEPT)
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...