Importance and interpretation of fast-response airway hygrometry during ventilation of anesthetized patients
Abraham Rosenbaum1, Peter H Breen
1Department of Anesthesiology, UCI Medical Center, University of California, Irvine, Building 53, Room 227, 101 The City Drive South, Orange, CA 92868, USA.
Background:
Measurement of oxygen uptake (Vo2) should help detect non-steady state critical events and metabolic derangement during anesthesia. Vo2 requires measurement of respiratory relative humidity (RH) and temperature (T). We have developed a fast response T and humidity sensor (HS), which uses tiny wet and dry thermometers to determine RH by psychrometry, where low RH causes evaporation to decrease wet T below dry T. In laboratory bench studies, we determined that >/=5 l/min gas flow through the HS is required for valid psychrometry function. This study demonstrates that monitoring of flow through the HS enhances the accuracy of RH measurement and interpretation.
Methods:
Phase One: Laboratory bench validation; We designed a special bench setup for the validation of metabolic gas exchange compared to precise ethanol combustion. Phase 2: Clinical study; During mechanical ventilation of 6 anesthetized surgical patients, airway flow was used to successfully select valid wet T and dry T during inspiration and expiration, from which respective RH's were calculated using principles of psychrometry.
Results:
The average (+/-SD) percent error for airway Vco2 (compared to the stoichiometric value) was -1.84 +/- 2.69% (Table 2). The average (+/-SD) percent error for airway Vo2 was 0.91 +/- 3.10%. Average RQ was 0.649 +/- 0.017. For all patients, average inspired RH was 36.1 +/- 11.8% (range of 17-52%), which differed significantly from expiration (103 +/- 9%). Among the 6-8 consecutive breaths for each patient, average standard deviations of expired RH were only 0.6%.
Conclusion:
We conclude that airway flow monitoring enhances the interpretation and accuracy of the fast-response HS measurements during inspiration and expiration, allowing for the determination of Vo2 in patients during anesthesia.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Airway, Skin Color, and Use of Accessory Muscles
Introduction
The initial evaluation of a patient's respiratory system...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Acute Respiratory Failure-IV