Related Experiment Video
Updated: Jul 7, 2026

Dual Test Gas Pulmonary Diffusing Capacity Measurement During Exercise in Humans Using the Single-Breath Method
Published on: February 2, 2024
Novel portable device measures preoperative patient metabolic gas exchange
Abraham Rosenbaum1, Heike C Howard, Peter H Breen
1Department of Anesthesiology, University of California, Irvine, California, USA.
Background:
Indirect calorimetry (IC), the measurement of airway CO2 elimination (VCO2), O2 [corrected] uptake (VO2) [corrected], and respiratory exchange ratio (RER = VCO2/VO2), is a noninvasive modality for the assessment of body metabolism. In anesthesia, IC can signal critical events and onset of acute metabolic derangements. We have previously demonstrated the accuracy and precision of a new IC measurement system designed for mechanically ventilated patients, comprised of a new clinical bymixer, fast response humidity and temperature sensor, and a flowmeter. However, measurement of IC during spontaneous breathing is challenging because of unstable tidal volume, frequency, and functional residual capacity (FRC).
Methods:
A new device for IC measurements, designed specifically for spontaneous breathing, was validated against a metabolic lung simulator bench setup. In a second study, the same device was used to conduct preoperative measurements of VCO2 and VO2 in 15 patients.
Results:
Our measurements showed excellent correlation and agreement with metabolic lung simulator values: The average (+/-SD) percent error for airway VCO2 was -4.7% +/- 3.31%; the average (+/-SD) percent error for airway VO2 was -0.30% +/- 5.25%. Average values of VCO2 and VO2 in the patient study (3.01 +/- 0.56 and 3.44 +/- 0.69 mL x kg(-1) x min(-1), respectively) were in agreement with previously reported values.
Conclusion:
We have shown that the new, portable bymixer-flow device, using a bymixer and a fast response humidity sensor, provided accurate and convenient bedside measurement of VCO2 and VO2. We believe that it can contribute in the future to preoperative assessment and baseline reference value for perioperative management.
More Related Videos
07:09Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like asthma or COPD,...
Gas Exchange and Transport