Multiplexed Fluorometric ImmunoAssay Testing Methodology and Troubleshooting
Published on: December 12, 2011
Variance in multiplex suspension array assays: carryover of microspheres between sample wells
1Microbiology Graduate Group, University of California, Davis, CA 95616, USA. bphanley@ucdavis.edu
Journal of Negative Results in Biomedicine
|April 27, 2007
Summary
Accidental bead carryover in suspended microarray instruments can lead to unpredictable results. Two types of carryover were observed: one predictable and one large, unpredictable type requiring multiple replicates or intraplex methods.
Area of Science:
- Biotechnology
- Assay Development
- Analytical Chemistry
Background:
- Accidental observation of bead carryover from unused wells in 96-well plates.
- Unexplained outliers in replicated wells have been previously noted.
- Carryover issues were observed on both older and new suspended microarray instruments.
Purpose of the Study:
- Investigate the nature and extent of bead carryover in suspended microarray instruments.
- Characterize different types of carryover observed during experiments.
- Inform researchers and biostatisticians about potential data integrity issues.
Main Methods:
- Utilized a multiple blank following well experiment design.
- Conducted experiments on two separate suspended microarray instruments.
- Analyzed bead counts from sequential wells to identify carryover patterns.
Main Results:
- Observed significant and unexpected bead carryover between wells.
- Identified two distinct carryover types: small/predictable and large/unpredictable.
- Unpredictable carryover did not consistently decline between wells and poses a significant challenge.
Conclusions:
- Bead carryover significantly impacts suspended microarray instrument data analysis.
- Biostatisticians must be aware of this phenomenon when designing experiments.
- Mitigation strategies include using multiple replicated samples or an intraplex method.
- Applicability to oligonucleotide-based assays remains undetermined.