Early pacemaker battery depletion caused by a current leak in the output circuitry: rectification not exchange
Markus Roos1, Richard Kobza, Paul Erne
1Division of Cardiology, Kantonsspital Luzern, Luzern, Switzerland.
Abstract:
Malfunction of a dual chamber system pacemaker due to a current leak in the output circuitry leading to a rapid decline in battery longevity is described. Serial pulse amplitude alterations in defined device configurations enabled location of the defect in the 2.5-V output voltage level; the current leak disappeared in the 5-V output voltage level, and recurred in the 7.5-V output voltage level which combines both 2.5-V and 5-V output circuitries. Reprogramming of pulse amplitudes rectified device malfunction. Circuitry reprogramming of pacemakers with faulty unexplained high battery current can circumvent early replacement in management of device malfunction.