Takagi Glare Tester CGT-1000 for contrast sensitivity and glare testing in normal individuals and cataract patients
1NH&MRC Centre for Clinical Eye Research, Department of Ophthalmology, Flinders Medical Centre and Flinders University, Bedford Park, South Australia, Australia. konrad.pesudovs@flinders.edu.au
Purpose:
To investigate the sensitivity and repeatability of the Takagi Contrast Glare Tester CGT-1000 in normal individuals and those with cataracts.
Methods:
A prospective observational study was performed. The Takagi Contrast Glare Tester measures contrast sensitivity (CS) at 6 target sizes and 13 contrast levels (2.00 to 0.34 logCS). Testing follows a method of descending limits paradigm with a single reversal determining threshold. The CGT-1000 was administered with and without glare in 95 eyes of 61 cataract patients and 13 controls. The percentage floor (unable to see the highest contrast) and ceiling (able to see the lowest contrast) effects and correlations between CS and cataract grades were determined. The repeatability was evaluated using Bland-Altman limits of agreement and expressed as the coefficient of repeatability (COR). Factor analysis was used to test for redundancy within the 6 spatial frequencies.
Results:
In normal individuals, a high rate of ceiling effect varying with target size was noted--for 6.3 degrees, 4.0 degrees, 2.5 degrees, 1.6 degrees, 1.0 degrees, 0.7 degrees, ceiling effects were 68%, 58%, 18%, 11%, 4%, 2%, respectively, for no glare, and 47%, 42%, 8%, 2%, 2%, 2%, respectively, with glare. In cataract patients, floor effects were noted--3%, 0%, 3%, 7%, 19%, 62%, respectively, for no glare, and 3%, 3%, 6%, 14%, 44%, 79%, respectively, with glare. Correlations with cataract grades ranged from 0.10 to 0.61, being best for nuclear cataract. Repeatabilities expressed as COR were +/- 0.11, +/- 0.14, +/- 0.28, +/- 0.38, +/- 0.38, +/- 0.47 logCS, respectively. All spatial frequencies loaded heavily on one factor, indicating no gain in information from testing multiple target sizes.
Conclusions:
Sensitivity to the presence of cataract was good, but ceiling effects in normal individuals and floor effects in cataract patients limit accuracy. Repeatability was poor, but could be improved by testing less spatial frequencies more rigorously.
More Related Videos
Related Concept Videos
Impact of Individuals on Individuals
Normal Stress
When a rod is under axial loading, the internal forces and corresponding stress are normal to the plane of the section, so it is termed normal stress. It's important to...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Normal Distribution
Variation: Normal Distribution, Range, and Standard Deviation
Normal and Shear Force