In Vitro Evaluation of The Effects Of Er,Cr:YSGG and Diode Lasers Used on Titanium Cylinder
Published on: June 6, 2025
The use of Er:YAG laser for cavity preparation: an SEM evaluation
Patricia Moreira Freitas1, Ricardo Scarparo Navarro, Juliana Almeida Barros
1Department of Restorative Dentistry, School of Dentistry, University of São Paulo, São Paulo, Brazil.
Objective:
The purpose of this study was to evaluate morphological changes in cavities prepared by the Er:YAG laser (2.94 mum) at different parameters of irradiation and by a diamond bur.
Experimental Design:
Cavities were prepared on 27 human molars (n = 3): G1, 15 Hz/160 mJ enamel/6 Hz/200 mJ dentin; G2, 15 Hz/180 mJ enamel/6 Hz/200 mJ dentin; G3, 15 Hz/160 mJ enamel/6 Hz/250 mJ dentin; G4, 15 Hz/180 mJ enamel/6 Hz/250 mJ dentin; G5, 15 Hz/180 mJ enamel/10 Hz/180 mJ dentin; G6, 15 Hz/160 mJ enamel/10 Hz/180 mJ dentin; G7, 15 Hz/160 mJ enamel/10 Hz/160 mJ dentin; G8, 15 Hz/180 mJ enamel/10 Hz/160 mJ dentin; G9, diamond bur. For SEM analysis, samples were fixed (2.5% glutaraldheyde, 12 h, 4 degrees C), dehydrated (25-100% ethanol), dried, and sputter-coated with gold.
Results:
Despite the changes on energy and repetition-rate settings, all laser-treated samples showed no evidence of thermal damage or signs of burning and melting. Er:YAG laser ablated dental hard tissues showed exposed enamel prisms, dentin surface without smear layer, and opened dentinal tubules.
Conclusion:
Different Er:YAG laser parameters were effective for ablation of hard tissues, creating an irregular and microretentive morphological pattern without hard tissue damage.