Related Experiment Video
Updated: Jul 13, 2026

Enhancing Electrode Location Assessment in Cochlear Implantation via Computed Tomography Image Fusion
Published on: January 17, 2025
Intracochlear assessment of electrode position after cochlear implant surgery by means of multislice computer
Gijs K A van Wermeskerken1, Mathias Prokop, Adriaan F van Olphen
1Department of Otorhinolaryngology, Head and Neck Surgery, University Medical Center-Utrecht, P.O. Box 85500, 3508 GA Utrecht, The Netherlands. gijs@vanwermeskerken.com
Abstract:
The development of electrode arrays, the past years, has focused on modiolus-hugging cochlear implant electrodes. Besides, atraumatic implantation of electrodes is of importance for the use in hearing preservation, in cases of combined electric and acoustic stimulation. Intracochlear positioning of the individual electrodes by means of multislice computer tomography (CT) has not yet been shown. In this study we formulated and tested a CT imaging protocol for postoperative scanning of the temporal bone in cochlear implant subjects. Both a fresh human temporal bone and a fresh human cadaver head were implanted with a cochlear implant. Multislice CT was performed for adequate depiction of the cochlear implant. All scans were analyzed on a viewing workstation. After mid-modiolar reconstruction we were able to identify the intracochlear electrode position relative to the scala tympani and scala vestibuli. This was possible in both the implanted isolated temporal bone and the fresh human cadaver head. The feasibility of imaging the electrode position of the cochlear implant within the intracochlear spaces is shown with multislice CT. An imaging protocol is suggested.