Related Experiment Video
Updated: Jul 12, 2026

09:13
Remote Magnetic Navigation for Accurate, Real-time Catheter Positioning and Ablation in Cardiac Electrophysiology Procedures
Published on: April 21, 2013
Remote magnetic navigation: human experience in pulmonary vein ablation
Luigi Di Biase1, Tamer S Fahmy, Dimpi Patel
1Section of Cardiac Electrophysiology and Pacing, Department of Cardiovascular Medicine, Cleveland Clinic, Cleveland, Ohio, USA.
Journal of the American College of Cardiology
|August 28, 2007
Summary
Remote magnetic navigation (MN) for atrial fibrillation (AF) ablation is feasible but current catheter technology limits effective lesion creation. This impacts the success rate for treating AF patients.
Area of Science:
- Cardiology
- Electrophysiology
- Medical Devices
Background:
- Novel magnetic navigation (MN) systems aim to standardize ablation procedures.
- Reducing reliance on operator skill is a key goal for complex cardiac interventions.
Purpose of the Study:
- To assess the feasibility and efficacy of remote MN and ablation for treating atrial fibrillation (AF).
- To evaluate the performance of the Niobe II remote magnetic system in AF ablation.
Main Methods:
- 45 consecutive AF patients underwent ablation using the Niobe II system.
- A stepwise approach included circumferential pulmonary vein ablation (CPVA) and pulmonary vein antrum isolation (PVAI).
- Remote navigation utilized coordinate or wand approaches; ablation endpoint was pulmonary vein electrical isolation.
Main Results:
- The wand approach achieved 100% site targeting, compared to 60% with the coordinate approach.
- Only 8% of pulmonary veins (PVs) were electrically isolated after initial ablation steps.
- Effective lesions were achieved in 23 patients with conventional catheters; recurrence rates were 22% for complete PVAI and 90% for incomplete PVAI.
Conclusions:
- Remote navigation with magnetic systems is a feasible technique for AF ablation.
- Current catheter technology is insufficient for creating effective lesions in most cases.
- This limitation significantly impacts the cure rate for atrial fibrillation patients.