Related Experiment Video
Updated: Jul 11, 2026

10:55
Detection of Anti-MDA5 Autoantibodies Using HeLa Cells and Immunocytochemistry with Light Microscopy
Published on: October 31, 2025
[Analytical interference caused by monoclonal immunoglobulin M with cryoglobulin effect: a case report]
1Laboratoire de biochimie, Hôpital d'instruction des armées Laveran, Marseille. manuelaoliver@cegetel.net
Annales De Biologie Clinique
|October 5, 2007
Summary
Monoclonal immunoglobulins can cause unexpected analytical interferences in laboratory tests. A case report highlights how a monoclonal immunoglobulin M with cryoglobulin effect led to false protein electrophoresis results due to precipitation.
Area of Science:
- Clinical Biochemistry
- Immunology
- Laboratory Medicine
Background:
- Monoclonal immunoglobulins are common in clinical biochemistry laboratories.
- These proteins can cause unpredictable analytical interferences, complicating test interpretation.
Observation:
- A case report details a monoclonal immunoglobulin M exhibiting a cryoglobulin effect.
- This specific immunoglobulin precipitated entirely during protein electrophoresis.
Findings:
- The precipitation of monoclonal immunoglobulin M resulted in a false-negative or inaccurate protein electrophoresis result.
- Cryoglobulin effects associated with monoclonal immunoglobulins are a significant source of analytical error.
Implications:
- Clinicians must be aware of potential interferences from paraproteins like monoclonal immunoglobulins.
- Collaboration between clinicians and laboratories is crucial when unexpected results arise, especially those potentially caused by monoclonal proteins.
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively coupled plasma–mass spectrometry (ICP–MS) is a highly selective and sensitive technique for accurate elemental analysis. Though the analysis of ICP–MS mass spectra is comparatively straightforward, it is affected by spectroscopic and non-spectroscopic interferences. Spectroscopic interferences arise when the plasma contains ionic species with an m/z value the same as the analyte ion. Spectroscopic interference can be categorized as isobaric, polyatomic ions, and refractory oxide ion...
Cross-reactivity
Overview