Sudden failure of implantable pulse generators: cause of failure and examination
Hermann Lanmüller1, Johanna Buchroithner, Johann Wernisch
1Department of Biomedical Engineering and Physics, Medical University of Vienna, Vienna, Austria. hermann.lanmueller@meduniwien.ac.at
Abstract:
A sudden failure of implantable pulse generators used for spinal cord stimulation occurred in two patients. To identify the cause of this failure, an intensive destructive analysis of the explanted devices was carried out. A functional diagnosis was carried out by inspecting amplitude, pulse width and frequency on each output channel of the implantable pulse generators. Later, the titanium case of the pulse generators was opened by laser cutting to minimise any additional mechanical stress during the opening procedure. The functional test for both pulse generators showed faultless behaviour. Using light and electron microscopy, hairline cracks could be identified in the electrical connection between battery and electronic circuit. In both devices, the cracks spread through the whole bond wire in the connection to the plus pole of the battery and partially also to the minus pole. The analysis showed that both devices failed by broken bond wires. The electrical connection to the battery exists just by the spring characteristic of the wires. A push to the implant causes a short-term disconnection, resulting in a power on reset of the device. Manufacturing or design issues, allowing micromotion between battery and the hybrid part, may be the reason for this problem.
Related Concept Videos
Special considerations while measuring pulse
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Cardiovascular System Abnormal Findings I: Inspection and Palpation
Abnormal findings observed during an inspection
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Pulse
The pulse serves as a clinical indicator...