Smart Bag vs. Standard bag in the temporary substitution of the mechanical ventilation
Robin Lovat1, Christine Watremez, Michel Van Dyck
1Cliniques Universitaires Saint Luc, Cardiovascular Department, Intensive Care Unit, Avenue Hippocrate 10, 1200 Brussels, Belgium. lovatrobin@yahoo.fr
Objective:
To compare in intubated patients manually ventilated in order to mirror the ventilator, the respiratory and hemodynamic effects induced by a bag device equipped with an inspiratory gas flow-limiting valve (Smart Bag, 0-Two Medical Technologies Inc., Mississauga, ON, Canada) and a Standard bag.
Design:
Non-randomized crossover study comparing 13 respiratory and eight hemodynamically paired parameters. Eight intubated patients were manually ventilated, each by three different intensive care workers yielding 24 sets of data for comparison. Data were collected during two sessions of manual ventilation, first with the Standard bag and then with the Smart Bag. Between each session, the patient was reconnected to the ventilator until return to the baseline. Patients, included after coronary surgery, were sedated and paralyzed.
Setting:
Intensive Care Unit, university hospital.
Results:
Compared with Standard bag, the Smart Bag provided a decrease of inspiratory flow (23 +/- 4.7 vs. 47.3 +/- 16.5 l/min) with a decrease of peak pressure (13.3 +/- 2.9 vs. 21.9 +/- 7.3 cmH2O) and tidal volume (9.4 +/- 2.8 vs. 12.4 +/- 2.7 ml/kg). While the expiratory time was similar, the inspiratory time increased (1.83 +/- 0.58 vs. 1.28 +/- 0.46 s) with the Smart Bag, limiting the respiratory rate (14 +/- 5 vs. 17 +/- 6 cycles/min) and the minute volume (8.8 +/- 2.9 vs. 14.4 +/- 4.9 l/min). Finally, it limited the fall of the ETCO2 (27.9 +/- 5.1 vs. 24.3 +/- 5.7 mmHg) and probably the risks of severe respiratory alkalosis. The bags similarly affected hemodynamic states.
Conclusion:
In intubated patients manually ventilated, the Smart Bag limits the risks of excessive airway pressure and the fall of the ETCO2, with hemodynamic effects similar to those of the Standard bag.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...