Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Automated Microbial Diagnostics01:24

Automated Microbial Diagnostics

Automated diagnostic analyzers have transformed clinical microbiology by providing rapid and reliable methods for pathogen identification and antibiotic susceptibility testing. Among these systems, the Vitek 2 is widely used because it automates the traditionally labor-intensive processes of microbial identification (ID) and antibiotic susceptibility testing (AST), delivering standardized and timely results that are essential for effective patient care.Microbial Identification with ID CardsThe...
Therapeutic Drug Monitoring: Drug Analysis Methods01:26

Therapeutic Drug Monitoring: Drug Analysis Methods

Therapeutic Drug Monitoring (TDM) is a clinical practice that measures specific drug levels in a patient's blood or body tissues to tailor drug therapy effectively. This monitoring is critical for managing drugs with narrow therapeutic indices like digoxin and phenytoin, ensuring they are both safe and effective. For instance, monitoring theophylline levels in asthma patients involves precision and sensitivity to adjust doses according to individual responses to therapy, ensuring efficacy and...
In Vitro Drug Release Testing: Overview, Development and Validation01:10

In Vitro Drug Release Testing: Overview, Development and Validation

In vitro dissolution and drug release tests assess how quickly and how much of a drug is released from its dosage form into an aqueous medium under standardized laboratory conditions. These tests are essential tools in pharmaceutical development and quality assurance, offering insight into the drug's performance before clinical use.During formulation development, dissolution testing identifies incomplete or inconsistent drug release issues. It also supports decisions on selecting the optimal...
Serum Laboratory Studies, Stool Test, Breath Test01:30

Serum Laboratory Studies, Stool Test, Breath Test

Gastrointestinal (GI) diagnostic studies are pivotal in confirming, ruling out, diagnosing, or staging various diseases, including cancers. Following diagnosis, allocating time for discussions with the patient and providing informational resources is crucial. Diagnostic assessments of the GI tract often occur in outpatient settings like endoscopy suites or GI labs. Preparation for these tests may include dietary restrictions, fasting, liquid bowel preparations, laxatives, enemas, and the...
Urine Studies I: Urinalysis01:29

Urine Studies I: Urinalysis

Urinalysis is a widely used diagnostic test that analyzes urine's physical, chemical, and microscopic characteristics. Healthcare providers use it to detect and monitor various health conditions, including renal disease, urinary tract infections (UTIs), diabetes, and metabolic or systemic disorders.Components of UrinalysisUrinalysis consists of three primary components: physical, chemical, and microscopic examination. Each provides unique insights into the urine sample and, by extension, the...
Effects of EDTA on End-Point Detection Methods01:18

Effects of EDTA on End-Point Detection Methods

Different methods, such as visual observance of metal-ion indicators, spectroscopic techniques, and potentiometric methods, can determine the endpoint of an EDTA titration.
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Clinical Utility of Whole RNA Sequencing for Fusion Detection in Acute Leukemia.

Cells·2026
Same author

Practical Considerations for Reporting <i>Staphylococcus epidermidis</i> Using BioFire FilmArray Blood Culture Identification 2 Panel: Lessons from Mixed Blood Cultures.

Annals of laboratory medicine·2026
Same author

Integrating Rapid Lymphoma Next-Generation Sequencing Panel-Based Testing into Routine Genomic Diagnostics.

Annals of laboratory medicine·2026
Same author

Comparative evaluation of two NGS-based assays for somatic hypermutation analysis of IGHV genes in chronic lymphocytic leukemia.

Blood research·2026
Same author

Germline Mutations Related to Complete Remission After Neoadjuvant Chemotherapy in Patients With Triple-negative Breast Cancer.

Journal of breast cancer·2026
Same author

Performance Evaluation of PacBio PureTarget for Multiple Short Tandem Repeat Expansion Detection.

Annals of laboratory medicine·2026

Related Experiment Video

Updated: Jul 9, 2026

Assessment of Ultrasonic Vocalizations During Drug Self-administration in Rats
09:25

Assessment of Ultrasonic Vocalizations During Drug Self-administration in Rats

Published on: July 22, 2010

[Evaluation of Viva-E drug testing system].

Hae-Sun Chung1, Seung-Tae Lee, Soo-Youn Lee

  • 1Department of Laboratory Medicine and Genetics, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea.

The Korean Journal of Laboratory Medicine
|December 21, 2007
PubMed
Summary

The Viva-E Drug Testing System demonstrates excellent analytical performance for therapeutic drug monitoring (TDM), offering a wide drug menu and efficient analysis. This makes it a valuable tool for clinical laboratories needing reliable drug level measurements.

More Related Videos

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
10:17

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry

Published on: April 23, 2019

Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
05:58

Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes

Published on: March 22, 2022

Related Experiment Videos

Last Updated: Jul 9, 2026

Assessment of Ultrasonic Vocalizations During Drug Self-administration in Rats
09:25

Assessment of Ultrasonic Vocalizations During Drug Self-administration in Rats

Published on: July 22, 2010

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
10:17

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry

Published on: April 23, 2019

Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
05:58

Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes

Published on: March 22, 2022

Area of Science:

  • Clinical Chemistry
  • Laboratory Medicine
  • Pharmacology

Background:

  • Therapeutic drug monitoring (TDM) is crucial for optimizing patient treatment.
  • Increased demand for drug analysis in clinical laboratories necessitates efficient instrumentation.
  • The Viva-E Drug Testing System (Dade Behring Inc.) is a new immunoassay instrument for TDM.

Purpose of the Study:

  • To evaluate the analytical performance of the Viva-E system.
  • To assess the clinical usefulness of the Viva-E system for TDM.
  • To compare Viva-E performance against established methods.

Main Methods:

  • Analytical performance (linearity, precision, correlation) assessed using patient samples and controls for 11 drugs.
  • Comparison with Cobas Integra 800 and API 4000 LC-MS/MS systems.
  • Analysis time was also evaluated.

Main Results:

  • Viva-E demonstrated good linearity (r2 ≥ 0.97) for all tested drugs.
  • Precision was acceptable (within-run CVs <5%, total CVs <10%) for most drugs.
  • Strong correlations (r=0.9283-0.9778) were observed with other methods.

Conclusions:

  • Viva-E exhibits strong analytical performance (linearity, precision, accuracy) for TDM.
  • Its comprehensive drug menu and efficient functions support its utility in clinical labs.