Troubleshooting FoCUS Image Acquisition: Patient Positioning, Transducer Manipulation, and Image Optimization
Published on: March 3, 2023
Optimizing imaging techniques in the postoperative patient
1Department of Radiology, Indiana University School of Medicine, University Hospital, Indianapolis, Indiana 46202, USA. kbuckwal@iupui.edu
Postoperative imaging with metal implants can cause image distortions. This article explains how to minimize these artifacts in computed tomography (CT) and magnetic resonance imaging (MRI) for better diagnostic accuracy.
Area of Science:
- Medical Imaging
- Radiology
- Biomedical Engineering
Background:
- Postoperative patients often require medical imaging to diagnose complications.
- Metal implants, common in postoperative patients, can significantly distort imaging results.
- Understanding and mitigating metal artifacts is crucial for accurate diagnosis.
Purpose of the Study:
- To explain the causes of metal artifacts in computed tomography (CT) and magnetic resonance imaging (MRI).
- To outline strategies for reducing or eliminating these artifacts in CT and MRI.
- To improve diagnostic accuracy in postoperative patients with metallic hardware.
Main Methods:
- Focuses on cross-sectional imaging techniques: CT and MRI.
- Analyzes factors influencing CT artifacts: hardware alloy, geometry, and location.
- Discusses artifact reduction techniques for CT: scan technique alteration, patient positioning, reconstruction algorithms, and multiplanar reformations.
- Examines factors influencing MRI artifacts: hardware alloy.
- Details artifact reduction techniques for MRI: pulse sequences (fast/turbo spin echo, inversion recovery), receiver bandwidth manipulation, and frequency encode axis orientation.
Main Results:
- CT artifacts are dependent on hardware alloy, geometry, and location.
- CT artifacts can be reduced by adjusting scan parameters, patient position, and reconstruction methods.
- MRI artifacts are primarily dependent on hardware alloy.
- MRI artifacts can be minimized using specific pulse sequences and modifying scan parameters like receiver bandwidth and frequency encode axis.
Conclusions:
- Metal artifacts in CT and MRI pose diagnostic challenges in postoperative patients.
- Specific technical adjustments in CT and MRI can effectively reduce or eliminate metal artifacts.
- Implementing these strategies enhances the reliability of diagnostic imaging in patients with metallic implants.
Related Concept Videos
Imaging Studies II: Ultrasonography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies VII: Vascular Imaging